設為首頁 | 加入收藏 | 聯系張經理:13608675847
發(fā)布者:本網 | 瀏覽次數: 次
Determination of Trace Manganese by Square-Wave Stripping
Voltammetry
Shujie Wang, Baoxian Ye*
Department of Chemistry, Zhengzhou University, Zhengzhou 450052, P. R. China
*e-mail: yebx@zzu.edu.cn
Received: August 6, 2007
Accepted: January 9, 2008
Abstract
A novel method of determining trace manganese by square-wave stripping voltammetry with simultaneous plating
mercury electrode is described in this paper. Well-defined stripping peaks of manganese are obtained in H3BO4-
NaOH buffer (pH 8.0) and peak currents are in good linear relationship with manganese concentrations in two ranges:
from 0.36 nM to 36 nM and from 73 nM to 909 nM. The obtained detection limit is 0.18 nM and relative standard
deviation is 2.3% (n?;;;7) under a predeposition time of 90 s. This proposed method has been used for the determination
of trace manganese in real plasma samples with satisfiable results and good matching with the results of graphite
furnace atomic absorption spectrometry.
Keywords: Manganese, Square-wave stripping voltammetry, H3BO4-NaOH buffer, Plasma samples